PureAir 100 & 50
Two sizes, one filtration chain.

Automatic particulate and toxic gas filtration system

An advanced air filtration system designed for laboratories and controlled environments.
How it works
Through its touchscreen interface, PureAir 200 provides real-time monitoring of temperature (°C), humidity (%), particulate levels and T-VOC (Total Volatile Organic Compounds).
Particulate sensor
T-VOC sensor
Temperature and humidity
Touchscreen
Real-time monitoring
Fan motor
4-mode speed
Inputs: particulate and T-VOC readings
Particulate sensor
T-VOC sensor
Temperature and humidity
Touchscreen
Real-time monitoring
Fan motor
4-mode speed
Inputs: particulate and T-VOC readings
Based on data received from the TVOC and particulate sensors, the fan speed is automatically optimized, ensuring that ambient conditions remain stable within the preset threshold values.
Sensors
PureAir 200 measures and displays particles of 0.3, 0.5, 1, and 5 µm and the total volatile organic compounds (T-VOC) in the environment in real time.
Air quality classification
Temperature (°C) and humidity (%) are monitored in real time through the touchscreen interface.
User-friendly interface
The 7″ capacitive, full-color TFT touchscreen provides easy and intuitive operation. It can be used comfortably with gloves and is resistant to liquid spills.

Screens are examples; the values shown are not measurement results.
Temperature (°C) and humidity (%) are monitored in real time.
The air quality classification for particles ranges from ISO CLASS-5 to ISO CLASS-9.
Readings for the four particle sizes are displayed in real time.
Total volatile organic compounds in the environment are displayed in real time.
The screen lists five classes, from ISO CLASS-5 to ISO CLASS-9.
Alongside the ISO classes, the screen also offers a manual option.
TVOC measurements can be displayed in ppm (parts per million) or ppb (parts per billion).
Particulate measurements are displayed in two units: particles/L and particles/m³.
Filters
PureAir 200 houses an advanced pre-filter assembly, activated carbon cartridge filters and a HEPA filter. Air enters the unit at the bottom and passes through the G4 pre-filters, the activated carbon cartridges and the H13 HEPA filter before leaving at the top.

Stage 1
Two ISO Coarse (G4) pre-filters with heights of 45 mm and 95 mm.
Stage 2
Four cylindrical activated carbon cartridge filters designed to target VOCs.
Stage 3
Equipped with a High-Efficiency Particulate Air (HEPA) filter of H13 grade.
Maintenance and alerts
PureAir 200, equipped with an integrated differential pressure sensor, provides both audible and visual alerts in the event of potential filter blockages.
The PureAir 200 alarm system monitors the optimal service life of your filters, automatically tracks replacement intervals, and notifies you when it is time to change them.
| Set | Codes |
|---|---|
| Semi-Annual Filter Set (6 Months) | PASA-00111RC048001 |
| Annual Filter Set (12 Months) | PAAF-11111RC048111 |
Fault diagnosis and repair, maintenance, filter replacement and installation.
PureAir 200
| Particle | T-VOC | |
|---|---|---|
| Sensor feature | Laser scattering | MOX-based digital gas sensing technology |
| Air quality classification | ISO CLASS-5, ISO CLASS-6, ISO CLASS-7, ISO CLASS-8, ISO CLASS-9 | 1 (Excellent), 2 (Good), 3 (Moderate), 4 (Poor), 5 (Unhealthy) |
| Units | Particles/m³ – Particles/L | ppm – ppb |
| Particle sizes | 0.3 µm, 0.5 µm, 1 µm, 5 µm | – |
Air Filtration Systems